The AC Hall Measurement Micro Probe System (ACHS) combines Nextron’s compact probe platform with AC Hall measurement technology, providing an integrated platform for characterizing electrical transport properties. By applying an alternating magnetic field and lock-in amplifier-based phasesensitive detection, the ACHS enables sensitive detection of small Hall signals while reducing the effects of DC offset and background noise, making it particularly suitable for low-mobility or high-resistance materials. Combined with Nextron’s environmental control technology, the ACHS enables Hall measurements under a wide range of temperature, vacuum, and gas conditions.

Min. air gap 29 mm
Magnet Type N52 Neodymium2)
Motorized Magnetic Rotation Speed Adjustable
Selectable Sample Contact Types Manual Probe / Motorized Probe / Wire Connection
Number of Probes (or Pins) 4 or 6
Temperature ranges ACHS-CH RT to 1000°C
ACHS-PT -40 to 200°C
ACHS-LN 80K to RT
Sample Stage Dimension ACHS-CH Ø 12.7 mm
ACHS-PT 19 × 19 mm
ACHS-LN 20 × 20 mm
Chamber External Dimension (excluding ports) (W × D × H)1) 117 × 80 × 25 mm
System Dimensions (W × D × H)1) ACHS-CH / ACHS-PT 250 × 320 × 380 mm
ACHS-LN 420 × 320 × 380 mm
Environmental control Vacuum tight < 1 × 10-4 mbar
Humidity control option 4 - 95 %RH
Vacuum partial pressure control option 0.001 - 1300 mbar
MFC Gas control option
Signal port SMA (SMA to BNC cable included)
Vacuum & gas port 1/4 in. & 3/8 in. tube fitting (Swagelok®-type)
Field Monitoring Resolution1 G
Electrical characteristics Leakage Current 100pA / 100 fA
Maximum DC Voltage, Current 300 VDC, 1000 mA
Frequency DC – 300 MHz

1) Specifications may vary depending on the temperature range, sample contact type, magnet type, vacuum configuration,
and sample size
2) For the ACHS-LN model with Wire Connection type.

(Unit: mm)
ACHS's dimension side view
Side View